Ion-Interaction Chromatography with Ion-Trap Mass Spectrometry Detection for the Analysis of Perchlorate in Drinking and Natural Waters

Main Article Content

Warawut Tiyapongpattana
Thanatcha Chaida
Napaporn Youngvises
Prapin Wilairat

Abstract

In this work, liquid chromatography with mass spectrometry (LC-MS/MS) for the determination of perchlorate in natural waters was developed. The method was based on ion-interaction (ion-pair) chromatography with the cationic ion-pairing reagent, hexamethonium bromide. Separation of the perchlorate was carried out using methanol and 1 mmol L-1 hexamethonium bromide (10:90, v:v) as the mobile phase at a flow rate of 0.40 mL min-1 and 100 µL injection volume. Under the optimal condition, the linearity range of perchlorate was 4 – 1000 µg L-1 with a coefficient of determination (r2) of 0.9998. The detection limit was 2 µg L-1, within U.S. EPA requirements (EPA methods 314.1 and 331.0). This method was effectively applied to the quantitation of perchlorate in drinking water and natural water samples.

Article Details

How to Cite
Tiyapongpattana, W., Chaida, T., Youngvises, N., & Wilairat, P. (2018). Ion-Interaction Chromatography with Ion-Trap Mass Spectrometry Detection for the Analysis of Perchlorate in Drinking and Natural Waters. Science & Technology Asia, 23(3), 35–43. Retrieved from https://ph02.tci-thaijo.org/index.php/SciTechAsia/article/view/147234